Electrosurgical Unit MEU-703

Description:

Electrosurgical Unit MEU-703 offers increased hemostasis through modulation frequencies of 20 kHz in Blend mode. For coagulation, its monopolar Forced Coag mode delivers a high voltage of 4800 V and a crest factor of 6.2 for rapid tissue desiccation. Its monopolar and bipolar modes provide flexibility across a wide range of clinical applications. Its different modes are tailored for fine surgical applications, minimizing collateral damage. Overall, this Electrosurgical Unit delivers reliable energy output, excellent tissue effects, and superior control for a wide variety of surgical needs.

Specifications :

Monopolar - Cut - Pure Cut Max Output Power 120W
Monopolar - Cut - Pure Cut Load Impedance 500Ω
Monopolar - Cut - Pure Cut Max Output Voltage 1300V
Monopolar - Cut - Pure Cut Crest Factor 1.8
Monopolar - Cut - Blend 1 Max Output Power 120W
Monopolar - Cut - Blend 1 Load Impedance 500Ω
Monopolar - Cut - Blend 1 Modulation Frequency 20kHz
Monopolar - Cut - Blend 1 Max Output Voltage 1400V
Monopolar - Cut - Blend 1 Crest Factor 2.0
Monopolar - Cut - Blend 2 Max Output Power 120W
Monopolar - Cut - Blend 2 Load Impedance 500Ω
Monopolar - Cut - Blend 2 Modulation Frequency 20kHz
Monopolar - Cut - Blend 2 Max Output Voltage 1300V
Monopolar - Cut - Blend 2 Crest Factor 2.0
Monopolar - Cut - Blend 3 Max Output Power 100W
Monopolar - Cut - Blend 3 Load Impedance 500Ω
Monopolar - Cut - Blend 3 Modulation Frequency 20kHz
Monopolar - Cut - Blend 3 Max Output Voltage 1400V
Monopolar - Cut - Blend 3 Crest Factor 1.9
Monopolar - Coag - Forced Max Output Power 120W
Monopolar - Coag - Forced Load Impedance 500Ω
Monopolar - Coag - Forced Modulation Frequency 25kHz
Monopolar - Coag - Forced Max Output Voltage 4800kHz
Monopolar - Coag - Forced Crest Factor 6.2
Monopolar - Coag – Soft Max Output Power 120W
Monopolar - Coag – Soft Load Impedance 500Ω
Monopolar - Coag – Soft Modulation Frequency 20kHz
Monopolar - Coag – Soft Max Output Voltage 1000V
Monopolar - Coag – Soft Crest Factor 2.0
Bipolar – Standard Max Output Power 100W
Bipolar – Standard Load Impedance 100Ω
Bipolar – Standard Modulation Frequency 20kHz
Bipolar – Standard Max Output Voltage 700V
Bipolar – Standard Crest Factor 1.9
Bipolar – Precise Max Output Power 50W
Bipolar – Precise Load Impedance 100Ω
Bipolar – Precise Modulation Frequency 20kHz
Bipolar – Precise Max Output Voltage 400V
Bipolar – Precise Crest Factor 1.9
Features :
    • Monopolar Cut Modes
    • Monopolar Coagulation Modes
    • Advanced Coagulation Modes
    • CQM (Contact Quality Monitoring) System
    • User Controls
    • Real-Time Parameter Display
    • Quick Adjustment
    • Intermittent Cut and Coagulation
  • Applications

    Our Electrosurgical Unit MEU-703 is designed for a wide range of surgical procedures across various specialties with its intelligent real-time output power feedback technology. This advanced system ensures precise energy delivery, excellent cutting performance, and minimal tissue damage.

    Standard Accessories :

    Accessories NameQuantity
    Monopolar Foot-Switch1 pc
    Ball4 pc
    Loop6 pc
    Blade2 pc
    Needle2 pc
    Triangle2pc
    Hand-Switch Pencil, Disposable1 pc
    Patient Return Electrode Without Cable, Split, for Adult, Disposable1 pc
    Connecting Cable for Patient Return Electrode (Split), 3m, Reusable1 pc
    Insulated Speculum with Smoke Tube1 pc