Ventilator System MVS-701 features pressure-trigger sensitivity ranging from -20 to 0 cmH₂O and configurable sigh functions and apnea ventilation settings. It is equipped with a 10.4-inch TFT screen for intuitive monitoring. This system supports multiple ventilation modes and manual ventilation to accommodate patient-specific needs. Its safety features include multiple alarms for oxygen concentration, airway pressure, minute volume, suffocation risk, and continuous pressure. With compact dimension and weight, it can be mounted on a trolley for mobility and ease of use in critical care settings.
| Ventilation Modes | IPPV, A/C, SIMV, SPONT/CPAP, SIGH, MANU |
|---|---|
| Ventilator Tidal Volume (Vt) | 20 mL to 1500 mL |
| Ventilator Frequency (Freq) | 1 min to 100 min |
| Ventilator I:E Ratio | 4:1 to 1:8 |
| Ventilator Oxygen Concentration | 45% to 100% |
| Ventilator PEEP | 0 cmH₂O to 20 cmH₂O |
| Ventilator Pressure Trigger Sensitivity (Ptr) | -20 cmH₂O to 0 cmH₂O (Based on PEEP) |
| Ventilator SIGH | 0 (off) / 1 / 100 to 5 / 100 |
| Apnea Ventilation | OFF, 5 s to 60 s |
| Monitoring Frequency (Freq) | 0/min to 100/min |
| Monitoring Tidal Volume (Vt) | 0 mL to 2000 mL |
| Monitoring MV (Minute Volume) | 0 L/min to 99 L/min |
| Monitoring Airway Pressure | 0 cmH₂O to 100 cmH₂O |
| Monitoring Oxygen Concentration | 15% to 100% |
| Low Battery Voltage Alarm | < 11.3 ± 0.3 V |
| High O₂ Concentration Alarm | 19% to 100% |
| Low O₂ Concentration Alarm | 18% to 99% |
| High Airway Pressure Alarm | 20 cmH₂O to 100 cmH₂O |
| Low Airway Pressure Alarm | 0 cmH₂O to 20 cmH₂O |
| High Minute Volume Alarm | 1 L/min to 99 L/min |
| Low Minute Volume Alarm | 0 L/min to 20 L/min |
| Continuous Pressure Alarm | PEEP + 1.5 kPa over 16 s |
| Suffocation Warning | 5 s to 60 s no spontaneous ventilation |
| Display | 10.4 |
| Pressure | 280 kPa to 600 kPa |
| Voltage | 220 V ± 22 V |
| Power Frequency | 50 Hz ± 1 Hz |
| Input Power | 40 VA |
| Oscillogram Display | P-T (Pressure–Time) F-T (Flow–Time) |
| Unit Dimensions (L × W × H) | 560 × 560 × 615 mm |
| Trolley Dimensions (L × W × H) | 650 × 680 × 620 mm |
| Unit Net Weight | 16 kg |
| Unit Gross Weight | 30 kg |
| Trolley Net Weight | 11 kg |
| Trolley Gross Weight | 37 kg |
Our Ventilator System MVS-701 is designed to provide advanced respiratory support across a wide range of clinical settings. It is ideal for use in intensive care units (ICUs), emergency departments, operating rooms, and during patient transport within healthcare facilities.